›› 2010, Vol. 41 ›› Issue (6): 905-908.doi: 10.3969/j.issn.0529-1356.2010.06.026

• 论著 • 上一篇    下一篇

腕关节多层螺旋CT扫描参数、图像质量和辐射剂量的关系

王向东1;韩萍1 ;赵虎2 ;柳曦1*   

  1. 1.华中科技大学同济医学院附属协和医院放射科; 2.华中科技大学同济医学院解剖学教研室,武汉 430022
  • 收稿日期:2009-08-21 修回日期:2009-11-12 出版日期:2010-12-06
  • 通讯作者: 柳曦

Correlation among scan parameter, image quality and radiation dose of cadaver wrist multislice spiral computed tomography

  1. 1.Department of Radiology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology;2.Teaching and Research Section of Aantomy,Tongji Medical College, Huazhong University of Science and Technology,Wuhan 430022, China
  • Received:2009-08-21 Revised:2009-11-12 Online:2010-12-06
  • Contact: LIU Xi

关键词: 腕关节, 扫描参数, 图像质量, 辐射剂量, 电子计算机体层成像,

Abstract: Objective To study correlations among scan parameter, image quality and radiation dose of cadaver wrist multislice spiral computed tomography(MSCT). Methods Four cadaver wrists labeled as 1-4 respectively underwent 16-slice CT (Siemens Somatom Sensation16) scan. Totally 11 groups of MSCT scan were performed for each specimen.Scan parameters were different. An additional scan was performed for the third specimen with a high voltage(140kV) and a high current(300mAs).Thin-slice coronal images were reformatted for all the scans.Each reformatted coronal image was evaluated. Correlations among scan parameter, image quality and radiation dose were analyzed. Results Image quality of each wrist with different scan parameters did not show evident difference. Radiation dose increased linearly with current and increased curvedly with voltage. Radiation dose did not change obviously when pitch was changed. Conclusion Relations are different between radiation dose and different scan parameters. Correlations among scan parameter, image quality and radiation dose are instructive in practice.

Key words: Wrist, Scan parameter, Image quality, Radiation dose, Computed tomography, Human