›› 2010, Vol. 41 ›› Issue (1): 157-159.doi: 10.3969/j.issn.0529-1356.2010.01.031

• 论著 • 上一篇    下一篇

利用数字散斑法测量股骨干骨折加压钢板内固定下的位移和形变

邹庆峰1;陈元晓2;李建云3;吴加权4;马琨4;陈英杰1*   

  1. 1.昆明医学院人体解剖教研室,昆明 650031;2.昆明医学院生物学教研室,昆明 650031;3.昆明理工大学力学系,昆明 650101;4. 昆明理工大学物理系,昆明 650101
  • 收稿日期:2008-10-04 修回日期:2008-12-30 出版日期:2010-02-06
  • 通讯作者: 陈英杰

Measuring the displacement and deformation under the internal fixation with compressing steel plate for femoral fractures by digital speckle method

  1. 1.Department of Anatomy, Kunming Medical University, Kunming650031, China;2.Department of Biology, Kunming Medical University, Kunming650031, China;3. Department of Mechanics, Kunming University of Science and Technology, Kunming650101, China;4. Department of Physics, Kunming University of Science and Technology, Kunming650101, China
  • Received:2008-10-04 Revised:2008-12-30 Online:2010-02-06
  • Contact: CHEN Ying-jie

关键词: 股骨干, 骨折, 钢板内固定, 生物力学,

Abstract: Objective Measuring the displacement and deformation of the fifth screw under different tensile force before or after adultal internal fixation with compressing steel plate for femoral fractuer by digital speckle correlation method.And analysis of the displacement and deformation after reducing the number of screws.Aimed at providing a theoretical basis for clinical femoral internal fixation with compressing steel plate about the number of screws. Methods Taking the antisepsis femoral specimen of grown-ups, embedding and fixing with tooth powder after removing the soft tissue, producing a fixed model in the 1/2 femur.Fixing steel plate below the lateral femur subperiosteal with five screws through the fourth floor of cortical bone on both sides of the fractures, samples were divided into 10 states, group A-J were measured by orders, under the rally of 400 N,500N , measuring the displayment of the fifth screw with CSS-44100-e-universal test machine for loading rally experiment ,and automatic data processing. Results For analysis of variance, the mean displacement measurements of the fifth screw under rally of 400N,500N, B-J group and A group have significant differences(EM>P/EM><0.05).The mean deformation measurements of the fifth screw - under rally of 400N,500N, group A and group F-J group have significant differences(EM>P/EM><0.05); group A and B-E group have no significant differences.Conclusion In a certain range, the fracture displacement and the numuber of the screws show negative correlation. The screws near the fracture line for femoral compression plate internal fixation of femoral fractuer is easily broken because

Key words: Shaft of femur, Fracture, Internal fixation with compressing steel plate, Biomechanics, Human

中图分类号: